Mass controls – RADWAG BALANCES 3Y SERIES User Manual

Page 180

Advertising
background image

- 180 -

Example no. 2 a balance AS 220.3Y:

No. Tare value

Minimum

sample

Means of operation

1

220.0000 g 0.5000 g The minimum sample refers to all net

weighments that are weighed in a

packaging with optional mass from full

measuring range of a balance (the <TARE>

button applied), as well as on weighing

samples without the packaging (the

<TARE> button is not applied).

Example no. 3 a balance AS 220.3Y:

No. Tare value

Minimum

sample

Means of operation

1

0.0000 g

0.2500 g The minimum sample refers to all net

weighments that are weighed without a

packaging (the <TARE> button is not

applied).

The software identifies this setting as the

minimum sample is valid only for weighing

samples without a packaging. If tarring

option is applied then the software disables

the pictogram informing on minimum

sample in use, and renders the weighment

as carried out with tare applied, that does

not have a minimum sample criterion

determined.

The user can preview entered data, but data editing is disabled.

29.15. Mass controls

(mode not available in standard version of the balance)

Each product control carried out on a balance is sent to a connected printer and

saved in the < Mass controls >database. Each control saved in the database

features its specific number set at the moment of its completion.

Mass control number format:

y y / M M / d d / H H / m m, where:

yy – year of control end,

MM – month of control end,

dd – day of control end,

HH – hour of control end,

mm – minute of control end.

Balance user can preview data on each control.

Advertising