19.1. Comparison of sample and (reference) mass standard

Comparison of samples with a mass standard can be carried out through:

−

Giving mass of a standard, using <

Give reference mass> soft key;

−

Accepting current mass loaded on the weighing pan as a reference

mass, using <

Set as 100%> soft key;

−

Selecting a product from database of products, where parameter mass

of a product entry is defined, using <

Database of Products> soft

key.

PROCEDURE 1 – manual declaration of reference mass

•

Press <

Give reference mass > soft key

•

In an opened window type in value of reference mass and accept it by

pressing <

> key

•

All weighed products are compared with the reference mass, and the

display indicates difference between weighed objects in [%].

PROCEDURE 2 – accepting currently loaded mass as reference (standard)

•

Place a sample on balance’s weighing pan

•

On stabilization of measurement result press <

Set as 100%> soft

key

•

The display indicates 100.000%, mass is accepted as reference, and

automatically saved in (reference) mass standard field.

•

Unload the sample from balance’s weighing pan

•

All samples weighed afterwards are compared with the reference mass,

and the display indicates the difference, expressed in [%], for each

weighed sample in relation to the reference mass.

PROCEDURE 3 – selecting a product from Database of Products

•

Press <

Database of Products> soft key, select a product entry to be

weighed from a list of displayed products

•

On selecting, the grey coloured workspace automatically updates its

content with data on Reference Mass

•

Mass of the weighed product is automatically inserted in field: reference

mass, which is activated by pressing <

> soft key

•

The display indicates 0.00 % (if the weighing pan is unloaded)

•

All samples weighed afterwards are compared with the reference mass,

and the display indicates the difference, expressed in [%], for each

weighed sample in relation to the reference mass.