Acquisition Modules and Platinum Firmware

Digitiser Configuration

Strong motion mode requires that the highest tap (the one with the fastest
sample rate) is set to 200 samples per second. The 200 sps data are passed

through a fast-response, divide-by-ten IIR filter in order to produce the
low-latency 20 sps data which are used for the calculations. This 20 sps

low-latency stream can be enabled for direct output: it's stream ID will have
the suffix 'E'.

Strong motion mode can be configured using buttons on the web page. If the
highest tap is configured for 200 samples per second operation, the following

is displayed:

If the highest tap is not set to 200 samples per second, an additional button is
displayed:

The

first reconfigures the taps so that the highest

tap is set to 200 samples per second and then enables strong motion mode. If
the highest tap is already set to 200 samples per second, the

button is active and, when clicked, enables strong

motion mode.

When in strong motion mode, this section of the screen displays as:

The “Transmit 20sps low-latency waveforms” check-box enables or disables

the 20 samples per second stream, which is used internally to provide the
data for the strong-motion calculations but can also be transmitted, recorded,

and used like any other output stream.

Clicking the

button cancels strong motion mode and

returns the CMG-DM24 to normal operation.

The thirty additional streams produced when in strong motion mode are

shown in the following table, where S represents the System ID and C
identifies the component ('Z' for vertical, 'N' for North/South or 'E' for
East/West).