Acquisition Modules and Platinum Firmware

Monitoring Operations

SAMPLES/SEC command (described in the DM24 manual). For
example, the command

dm24­upgrade ­­samp 1000 500 100 50 PortA

will set tap 0 to 1,000 samples per second and taps 1, 2 and 3 to 500,
100 and 50 samples per second respectively.

­­cont # # # #

This option specifies which streams will be enabled for continuous
output at each tap. This argument requires requires four numerical

values to be supplied using the same format as for the CONTINUOUS
command (described in the DM24 manual). For example, the

command

dm24­upgrade ­­cont 1 0 0 7

will turn on continuous output of the vertical component at tap 0

and of the three triaxial components at tap 3

­­trig # # # #

This option specifies which streams will be enabled for triggered

output at each tap. This argument requires requires four numerical
values to be supplied using the same format as for the TRIGGERED

command (described in the DM24 manual).

­­gps­baud #

This option takes a single numerical argument which specifies the

Baud rate of the GPS port.

­­in­baud #

This option takes a single numerical argument which specifies the

Baud rate of the Data in port.

­­upgrade

Upgrade firmware files only (this is the default).

­­downgrade

This option over-rides internal version number checks, allowing
firmware to be downgraded as well as upgraded.

­­force

This option causes firmware files are to be loaded even if they appear
identical to the versions already loaded, or in cases where the

installed version number cannot be decoded.