Chapter 5 To Trigger the Oscilloscope

5-22

MSO2000A/DS2000A User’s Guide

UpLevel - LowLevel

SlewRate =

Time

When the Slope Condition is set to

or

, the current trigger level

and slew rate range will be displayed at the lower left corner of the screen, as shown
in figure 5-5 (b). The formula of slew rate range is:

UpLevel - LowLevel

UpLevel - LowLevel

SlewRate =

~

UpperLimit

LowerLimit

(a)

(b)

Note:

Under the “Slope” trigger menu, you can also press down the trigger level

knob continuously to switch the vertical window.

During the adjustment, two orange trigger level lines and two trigger marks ( and

) appear on the screen and move up and down with the rotation of the knob,

when stopping turning the knob, the trigger level lines and trigger marks disappear
in about 2 s.

Trigger Mode:

Press Sweep and select “Auto”, “Normal” or ”Single”. For the details, please refer to

Trigger Mode. The corresponding status light of the current trigger mode turns

on.

Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.