Time base mode, Y-t mode, X-y mode – RIGOL MSO/DS2000A Series User Manual

Page 70: Time base mode -4, Y-t mode -4 x-y mode -4

Advertising
background image

RIGOL

Chapter 3 To Set the Horizontal System

3-4

MSO2000A/DS2000A User’s Guide

Time Base Mode


Press MENU in the horizontal control area (HORIZONTAL) at the front panel and
then press Time Base to select the time base mode of the oscilloscope and the
default is Y-T.

Y-T Mode

In this mode, the Y axis represents voltage and the X axis represents time.

Note:

Only when this mode is enabled can Delayed Sweep be turned on. In this

mode, when the horizontal time base is greater than or equal to 200 ms, the
instrument enters slow sweep mode. For the details, please refer to the introduction
of slow sweep in Roll Mode.

X-Y Mode

In this mode, the oscilloscope changes the two channels from voltage-time display
mode to voltage-voltage display mode. Wherein, the X axis and Y axis tracks the
voltages of CH1 and CH2 respectively. The phase deviation between two signals with
the same frequency can be easily measured via Lissajous method. The figure below
shows the measurement schematic diagram of the phase deviation.












Figure 3-2 Measurement Schematic Diagram of Phase Deviation

II I

III IV

A

B

C

D

The signal must be
centered horizontally

Advertising