To set the waveform display size, Reorder setting, To set the threshold – RIGOL MSO/DS2000A Series User Manual

Page 175

Advertising
background image

Chapter 7 Digital Channel

RIGOL

MSO2000A/DS2000A User’s Guide

7-5

To Set the Waveform Display Size


Press LA WaveSize, rotate

to select the desired display size and the press

down

. You can also press WaveSize continuously to switch among S (small), M

(medium) and L (large).

Note:

L (large) can only be used when the number of channels currently turned on is

no more than 8.

Reorder Setting


Press LA ReOrder to select the waveform order mode on the screen of the
channels currently selected. You can select “D0-D15” or “D15-D0” and the default is
“D0-D15”.
 D0-D15: the waveforms on the screen are D0-D15 from top to bottom.
 D15-D0: the waveforms on the screen are D15-D0 from top to bottom.

To Set the Threshold


Press LA Threshold to enter the threshold setting menu. The threshold levels of
channels D7-D0 and channels D15-D8 can be set separately. When the voltage of the
input signal is greater than the threshold currently set, it is treated as logic 1;
otherwise, it is treated as logic 0.

Press Type to open the threshold type selection list to select a preset value or user.
 The preset values include TTL, 5.0V CMOS, 3.3V CMOS, 2.5V CMOS, 1.8V CMOS,

ECL, PECL, LVDS and 0V. After selecting a preset value, press AppToD7-D0 to
apply the threshold level to channels D7-D0 and press AppToD15-D8 to apply
the threshold level to channels D15-D8.

 When “User Def” is selected, press D7-D0 or D15-D8 and use

to set the

desired thresholds respectively. The range is from -20.0V to +20.0V and the
default is 700mV.

Advertising