Time parameters – RIGOL MSO/DS2000A Series User Manual

Page 150

Advertising
background image

RIGOL

Chapter 6 MATH and Measurements

6-14

MSO2000A/DS2000A User’s Guide

Time Parameters




Figure 6-3 Time Parameters


1. Period: defined as the time between the middle threshold points of two

consecutive, like-polarity edges.

2. Frequency: defined as the reciprocal of period.
3. Rise Time:
the time for the signal amplitude to rise from the threshold lower

limit to the threshold upper limit.

4. Fall Time: the time for the signal amplitude to fall from the threshold upper

limit to the threshold lower limit.

5. + Width: the time difference between the threshold middle value of a rising

edge and the threshold middle value of the next falling edge of the pulse.

6. - Width: the time difference between the threshold middle value of a falling

edge and the threshold middle value of the next rising edge of the pulse.

7. + Duty: the ratio of the positive pulse width to the period.
8. - Duty:
the ratio of the negative pulse width to the period.

Note:

The default values of the threshold upper limit, threshold middle value and

threshold lower limit in the figure above are 90%, 50% and 10% respectively. You
can set them via MeasureSettingType  “Threshold” and for the setting
method, please refer to the introduction in Threshold Measurement Setting.

RiseTime

FallTime

Threshold

Upper Limit

Threshold

Middle Value

Threshold

Lower Limit

+Width

-Width

Period

Advertising