Chapter 10 Pass/Fail Test

10-4

MSO2000A/DS2000A User’s Guide

waveforms are detected. The test results on the screen will update
continuously and a pulse will be output from the [Trigger Out] connector
at the rear panel each time a failed waveform is detected.


3. Beeper on Fail

Press Output to select “Fail” or “Fail+ ”.
 Fail: when a failed waveform is detected, the beeper does not sound.
 Fail+

: when a failed waveform is detected, the beeper sounds (not

related to the on/off state of the sound).


4. [Trigger Out] Outputs Pulse

Press AuxOutput to select “ON” or “OFF”. If “ON” is selected, the oscilloscope
outputs a pulse from the [Trigger Out] connector at the rear panel when a
failed waveform is detected.
You can also press UtilityAuxOutput and select “PassFail” to set this
output.

To Save the Test Mask


Users can save the current test mask in “*.pf” format to the internal Flash memory or
external USB storage device. The internal memory can store at most 10 test mask
files (LocalPF.pf).

Press Save to enter the file store interface. Please refer to the relative descriptions in

Store and Recall to save the test mask file to the internal or external memory.

To Load the Test Mask


Users can also load the test mask files (*.pf) stored in the internal Flash memory or
external USB storage device to the instrument.

Press Load to enter the file recall interface. Please refer to the relative descriptions
in

Store and Recall to load the test masks to the instrument.