General Specifications

Probe Compensation Output
Output Voltage

[1]

About 3 V, peak-peak

Frequency

[1]

1 kHz

Power
Power Voltage

100 V to 240 V, 45 Hz to 440 Hz

Power

Maximum 50 W

Fuse

2 A, T Degree, 250 V

Environment
Temperature Range

Operating: 0 ℃ to +50 ℃
Non-operating: -40 ℃ to +70 ℃

Cooling Method

Fan cooling

Humidity Range

0℃ to +30℃: ≤95% Relative Humidity
+30℃ to +40℃: ≤75% Relative Humidity
+40℃ to +50℃: ≤45% Relative Humidity

Altitude

Operating: under 3,000 meters
Non-operating: under 15,000 meters

Physical Characteristics
Size

[3]

Width×Height×Depth = 361.6 mm× 179.6 mm×130.8
mm

Weight

[4]

Package Excluded

3.9 kg ± 0.5 kg

Package Included

4.5 kg ± 0.5 kg

Calibration Interval
The recommended calibration interval is one year.
Regulatory Information
Electromagnetic
Compatibility

2004/108/EC
Execution standard EN 61326-1:2006 EN 61326-2-1:2006

Safety

UL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004;
EN 61010-1:2001; IEC 61010-1:2001

Note

[1]

: Typical.

Note

[2]

: Maximum value. 20ns, single-channel mode, dots display, auto memory depth.

Note

[3]

: Supporting legs and handle folded, knob height included.

Note

[4]

: Standard configuration.