Chapter 2 To Set the Vertical System

RIGOL

MSO2000A/DS2000A User’s Guide

2-5

Input Impedance


This oscilloscope provides two input impedance modes (1 MΩ (default) and 50 Ω) to

reduce the circuit load caused by the interaction of the oscilloscope and the circuit to

be tested.
 1MΩ: the input impedance of the oscilloscope is rather high and the current

flowing into the oscilloscope from the circuit under test can be ignored.

 50Ω: match the oscilloscope with a device with 50 Ω output impedance.

Press CH1Input to set the input impedance of the oscilloscope. When “50Ω” is
selected, “Ω” is displayed in the channel status label at the bottom of the screen.

Note:

The setting of the input impedance affects the settable range of the vertical

scale and vertical offset of the corresponding channel. For the details, please refer to
Specifications.

Waveform Invert


Press CH1Invert to enable or disable waveform invert. When waveform invert is
enabled, the displayed waveform rotates 180 degree relative to the ground potential.
When waveform invert is disabled, the waveform display is normal.

(a) “Invert” is disabled

(b) “Invert” is eabled

Figure 2-1 Waveform Invert