Delay calibration of the analog channel, Delay calibration of the analog channel -8 – RIGOL MSO/DS2000A Series User Manual

Page 64

Advertising
background image

RIGOL

Chapter 2 To Set the Vertical System

2-8

MSO2000A/DS2000A User’s Guide


Figure 2-2 Label Modification Interface


After finishing the input, press OK to finish the modification and the channel label
will change to “

”.


To modify or delete the input character, press Name to select the “Name Input Area”
and use

to select the character to be modified or deleted. Enter the desired

character to modify the character or press Delete to delete the character selected.

Delay Calibration of the Analog Channel


When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring relatively greater error (zero offset). Zero offset is defined as
the offset of the crossing point of the waveform and trigger level line relative to the
trigger position, as shown in the figure below.

Figure 2-3 Zero Offset

Name Input Area Keyboard Upper/Lower Case Switch