Chapter 5 To Trigger the Oscilloscope

RIGOL

MSO2000A/DS2000A User’s Guide

5-9

Edge Trigger


Trigger on the trigger threshold of the specified edge of the input signal.

Trigger Type:
Press Type, rotate

to select “Edge” and press down

. At this point, the trigger

setting information is displayed at the upper right corner of the screen. For example,

. The trigger type is edge trigger; the trigger source is CH1; the

trigger level is 0.00 V.

Source Selection:

Press Source and select CH1, CH2, EXT, AC Line or any channel of D0-D15. For the

details, please refer to the introduction in Trigger Source. The current trigger

source is displayed at the upper right corner of the screen.

Note:

Select channel with signal input as trigger source to obtain stable trigger.


Edge Type:
Press Slope to select the kind of edge of the input signal on which the oscilloscope

triggers. The current edge type is displayed at the upper right corner of the screen.

: trigger on the rising edge of the input signal when the voltage level meets

the preset trigger level.

: trigger on the falling edge of the input signal when the voltage level meets

the preset trigger level.

: trigger on the rising or falling edge of the input signal when the voltage

level meets the preset trigger level.


Trigger Mode:

Press Sweep and select “Auto”, “Normal” or ”Single”. For the details, please refer to

Trigger Mode. The corresponding status light of the current trigger mode turns

on.

Trigger Setting:

Press Setting to set the trigger parameters (trigger coupling, trigger holdoff and

noise rejection) under this trigger type.

Note:

Trigger coupling and noise rejection are grayed out and disabled automatically

when the source is set to any channel of D0-D15.


Trigger Level:
Trigger occurs only when the signal reaches the preset trigger level. Use TRIGGER