Chapter 1 Quick Start

1-12

MSO2000A/DS2000A User’s Guide

(2) Insert a small straight screwdriver into the slot at the power socket and pry

out the fuse seat gently.

(3) Take out the fuse and replace it with a fuse of the specified specification.

Then, reinstall the fuse seat.

5 AC Power Socket

AC power input terminal. The power requirement of this oscilloscope is 100-240
V, 45-440 Hz. Use the power cord provided with the accessories to connect the
instrument to AC power. Then, you can press the power key at the front panel to
start the instrument.

6 Lock Hole

You can lock the instrument to a fixed location using the security lock (please
buy it yourself) via the lock hole.

7 Trigger Out and Pass/Fail

Trigger Out

This connector will output a signal that can reflect the current capture rate
of the oscilloscope at each trigger. Connect the signal to the waveform
display device and measure the frequency of the signal. The measurement
result is the same with the current capture rate.

Pass/Fail

During the pass/fail test, this connector will output a positive pulse each
time a failed waveform is detected and low level when no failed waveform is
deteced.

8 Source Output Connectors

The output terminals of the built-in dual-channel source of the oscilloscope.
When Source1 or Source2 is enabled, the signal currently set can be output
through the [Source1] or [Source2] connector at the rear panel. This function
is only available for MSO2000A-S and DS2000A-S models oscilloscopes.