Chapter 9

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 105 of 199

During a CALIBRATION TASK the Monitor uses these Optical Filter

Factors to calculate the calibration factors for each installed optical

filter. As explained in

section 9.1.2

, by “entering” a Filter Bank No. in

the active set-up before calibration of any particular filter with any

particular gas, each filter can be calibrated to measure up to five

different gases and the filter bank number tells the Monitor exactly

where (in the Source Memory) to store the calibration factors

calculated during the calibration task.

Fig.9.2 The state of the Source Memory after completion of a

CALIBRATION TASK

After completion of a calibration task the calibration factors, which

are found in the Working Memory, are entirely dependent upon the

Filter Bank No., which is active for each optical filter. If Filter Bank

No. 1 is active for filters in positions “A” to “D” and Filter Bank No. 2

is active for the filter in position “E” then calibration factors from

files A 1, B 1, C 1, D 1, E 2 and W will be found in Working Memory

(see Fig 9.2).