Type of Span Calibration – Single-point or Two-point?

Fig.12.15a Graph showing a non- linear span-calibration curve which can

be plotted using the signals measured during a two-point span

calibration task

Fig.12.15b Graph showing a non- linear measurement curve

The question about whether to perform a single- or a two-point span

calibration is difficult to answer, as it is dependent on how linear the

relationship is between a gas’s concentration and the signal it

produces in the cell (see

Fig.12.14

,

Fig.12.15a

and

Fig.12.15b

).

However, we can give you some guidelines on whether you have to

perform a single- or a two-point span-calibration.