After scrolling through the list of individual measurement results for

gas A, the following text is displayed:

SAMPLING IS CONTINUOUS

MONITORING PERIOD IS NOT PRE-SET

COMPENSATION FOR WATER VAP. INTERFERENCE

MONITORING TASK STARTED 2014-01-16 09:50

MEASUREMENT CHAMBER TEMPERATURE 29.6°C

SAMPLE INTEGRATION TIME W: 5s

A: 5s B:___s C:___s D:___s E:___s

A:Carbon monoxide B:__________ C:___________

D: ___________ E:__________ W: WATER VAP

A: 7.42E+00 B: ________ C:___________

D: ____________ E:

________ W: 8.13+00

The last display above shows the up-dated overview of the gas

concentrations measured during the most recent measurement cycle.

This display does not appear, after the monitoring task has been

stopped.

Data collected during a monitoring task i stored in the Display

Memory of the Monitor and therefore, before a new monitoring task

is started, all data stored in this memory is deleted. Display Memory

has finite size, and therefore there is a limit to how much data can

be stored in it. Once this memory has been filled-up all subsequent

data collected during the task will start to overwrite the data already

collected during the start of the monitoring task. Only in this event

will the date displayed after the text MONITORING TASK STARTED

on the above screen be different from the earliest date on the

display (shown below) which give the most up-to-date overview of

the monitoring task: