LumaSense Technologies INNOVA 1314i User Manual

Page 81

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 81 of 199

Monitoring Task branch

2.

Press S1 to enter the Monitoring Task branch.

The following text is displayed:

3.

Up to 10 different monitoring tasks can be defined here. These

are labelled from 1 to 10. Use the direction push-buttons to

select the desired set-up number and move on to the next

display.

4.

Use the select push-buttons to define if you want to sample

continuously:

If you select No (S1), go to step 5.

If you select Yes (S3), go to step 6.

5.

The sampling interval is displayed. Use the direction push-

buttons to change it to the correct value.

6.

Use the select push-buttons to define if you want to monitor for

a pre-set period:

If you select Yes, go to step 7.

If you select No, go to step 8.

7.

The monitoring period is displayed. Use the direction push-

buttons to change it to the correct value.

The MONITORING PERIOD is entered as Days Hours: Minutes.

For example, if this particular measurement cycle needs to be

performed during 5 days, 14 hours and 30 minutes, the

“monitoring period” should be “entered” as 5 14:30. The Monitor

will then automatically stop monitoring 5 days, 14 hours and 30

minutes after the measurement cycle with this particular set-up

number is started. The range of acceptable values and the default

value of the monitoring period is shown in

Table 7.7

.

MONITORING TASK SET-UP NUMBER 1

PRESS ENTER TO CHANGE VALUE