Chapter 14

______________________________________________________________________

_____________________________________________________________________________

BE6030-12

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 190 of 199

If there is a perfectly linear relationship between the points C

low

, V

low

and C

high

, V

high

then:

V

lin

= ( C

h

/ C

l

) * V

l

GAS X: MEAS. INDICATES LINEAR COHERENCE

This message will be given in the status display if V

l

lies between the

values V

lin

– δ and V

lin

+ δ. This indicates that the two gas

concentrations you have chosen lie in the linear region of the

calibration curve. If the high concentration of gas you used (C

high

) is

equal to or greater than the highest concentration you expect to

measure, then it is not necessary to use a two-point span calibration,

a single-point span calibration using C

high

as the concentration of span

gas used will be sufficient to give you a good calibration result.

However, if the high concentration of gas you used is less than the

highest gas concentration you expect to measure, the two-point span

calibration task should be repeated using a high concentration which is

equal to or greater than the highest concentration you expect to

measure. There will therefore be a better chance to find a point in the

non-linear region of the span calibration curve so that any non-

linearity can be plotted.

GAS X: MEAS. DOES NOT FIT NON-LINEAR CURVE

This message will be given in the status display if V

l

is less than V

lin

δ. This message indicates that a two-point span calibration cannot be

performed.

If V

l

is found to be greater than V

lin

+ δ then the Monitor will try to fit

the two measured points onto a curve, whose shape has been pre-

determined by LumaSense, by an iterative process. The iterative

process involves scaling on the concentration and signal axes until the

two points fit the curve within certain tolerances.

GAS X: CALIBRATION IMPOSSIBLE: MAX ITER.

If after 255 iterative processes, the two points cannot be fitted onto

the curve within tolerance limits, this message is displayed.

This could indicate that the wrong concentration has been “entered” in

the set-up of the span calibration task (which does not relate to the

measured signal). Alternatively, it could indicate that there is a linear

relationship between the gas signal in the cell and the concentration of

the gas (in which case only a single-point span calibration is required).