2 symbols used in displayed measurement data, Section 8.2.2, Gas x: common mark: peow gas mark: bfa time: hh:mm – LumaSense Technologies INNOVA 1314i User Manual

Page 95: Common marks

Advertising
background image

Chapter 8

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 95 of 199

Where: N =

measurement cycles performed during the

monitoring period (T) between the cursors.

c

n

= gas concentrations (where n = 1 to N)

The Standard Deviation (σ) – this is the standard deviation of a

particular gas’s measurements from the mean value (µ) of this

gas. Standard deviation can be expressed mathematically as

follows:

The Standard Deviation =

(

)

1

1

2

-

-

=

å

=

=

N

Cn

N

n

n

m

s

Where: µ =

the mean value

C

n

=

individual gas concentrations

N

=

total number of measurements

8.2.2

Symbols used in Displayed Measurement Data

If an asterisk (*) appears on the display alongside any gas

concentration, it indicates that something happened during the

measurement. You can find out what the asterisk indicates by

pressing the INFO button. The following display appears for a short

period of time:

GAS X: COMMON MARK: PEOW GAS MARK: BFA

TIME: hh:mm

The symbols which can appear after the text COMMON MARK and

GAS MARK are listed below with a short explanation of what they

mean:

Common Marks:

“P”

(1) The Monitor stopped measuring due to a power

failure but has automatically re-started the

monitoring task again after regaining power;

(2) The Monitor has been “reset” either because of

a detected fault or because RESET was pressed.

“O”

This symbol indicates that an operational error was

detected during this measurement. The accuracy of

such a measurement cannot always be guaranteed.

Advertising