Fig.11.7 Refitting the retaining disc

11.1.2 Changing the Filter-paper in the External Air-

filtration Unit

While you do this, we recommend that you wear clean, rubber

gloves.

Tools and equipment required:

Acetone (analytically pure)

Spare fine filter-paper DS 0759

Tweezers with Teflon coating

Cotton buds

Clean rubber gloves

To change the filter-paper:

Switch off the Monitor.

Put a rubber glove on the hand which is to hold the parts of the unit.

Pull the external fine air-filtration unit off the end of the Teflon

®

sampling tube. Hold the stub-end of the external air-filter unit

between your thumb and fingers, and unscrew the coarse air filter

from the end of the unit. Refer to

Fig.11.3

.

Remove the used (old) filter-paper and the retaining disc from the

unit using the tweezers. Refer to Fig.11.8.

Fig.11.8 The parts of the external air-filtration unit UD 5023