Page 191

Advertising
background image

Chapter 14

______________________________________________________________________

_____________________________________________________________________________

BE6030-12

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 191 of 199

SPAN CALIBRATION IMPOSSIBLE

This message is associated with the span calibration of the water-

vapour filter.

The above message indicates that the Monitor cannot calculate the

gas conversion factor. This could indicate that the concentration of

water-vapour used during the calibration was too low – this could

either be due to an incorrectly “entered” water-vapour concentration

in the set-up of the span calibration or that the concentration of

water-vapour used was not the same as that which was “entered” in

the set-up.

14.4.2 “Successful*” Error Messages

WATER VAP. MEAS. < OFFSET MEAS

This message is associated with one, or more, of the following

calibration tasks: (1) the one- or two-point span calibration of an

optical filter; (2) the cross-interference calibration of the installed

optical filters ("A” to “E”); (3) the zero-point calibration of all the

installed optical filters (“A” to “E”).

During all the above-listed calibration tasks, the signal in the cell is

measured using the water-vapour filter: this enables the Monitor to

compensate for any interference signal produced by the presence of

any water-vapour in the span gas being used. If the signal in the cell

using the water-vapour filter is found to be less than the concen-

tration offset factor for the water-vapour filter, the above message

is sent.

This could indicate that the zero gas you used, during the zero-point

calibration of the water-vapour filter, was not perfectly dry.

SPAN MEASUREMENT TOO LOW

This message is associated with the span calibration of one (or more)

of the filters.

The above message indicates that the signal measured in the cell

during span calibration is less than 3µV and therefore cannot be used

to obtain a good calibration. This means that the concentration of your

span gas is too low.

SIGNIFICANT CHANGE IN CALIB. FACTORS

This message is associated with the span calibration of one (or more)

of the filters.