Chapter 4

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 33 of 199

Fig.4.1 An illustration of the fixed sampling intervals of the monitor

when it is not sampling continuously.

4.1.4

Continuous Sampling

The Monitor is said to be sampling continuously when each

measurement cycle is followed immediately by another similar

measurement cycle (see Fig.4.2).

Fig.4.2 An illustration of continuous sampling

4.1.5

Monitoring Period

The monitoring period is the time between the start of the first

measurement cycle in a monitoring task and the finish of the final

measurement cycle in a monitoring task.

The monitoring period can either be pre-set by the user – in which

case the Monitor automatically stops monitoring after the monitoring

period is over – or, not pre-set – in which case the Monitor will

continue to monitor until the user stops the monitoring task manually

(see

section 4.3.8

).The maximum time you can set a pre-set

monitoring period to be is 7 days. If You try to enter a greater

number of days, the Monitor will automatically change the period to

7 days. However, if you do not set a predefined monitoring period,