functioning correctly, it has not been span calibrated to

measure the gas you wish to measure.

2. Under the heading GENERAL INFORMATION FOR WATER

FILTER – ACTIVE CALIBRATION DATA : check the following

information:

a.

Zero-point Calibration performed: if a date appears here

this is the last time the water-vapour filter was zero-point

calibrated.

b.

Concentration offset factor: if a number appears here

this means that the water-vapour filter has been zero-point

calibrated.

c.

Span Calibration performed: if a date (year-month-

day) appears here, this is the last time span calibration of

the water-vapour filter was performed.

d.

Conversion Factor: if a number appears here this means

that the water-vapour filter has been span calibrated.

When the water-vapour filter has been zero-point calibrated and

span calibrated it is able to measure the absolute concentration of

water vapour in any gas sample in the analysis cell of the Monitor,

and it requires no further calibration.

When all installed optical filters (in positions “A” to “E”) have been

zero-point, humidity-interference, span and cross-interference

calibrated (if necessary) then the Monitor is able to perform accurate

measurements.