12.5.1 Retaining Existing Calibration Factors

When zero-point, humidity-interference and cross-interference

calibrations are performed, all the installed filters are calibrated. In

some case, you may wish to keep the calibration factors already

present in your monitor. If this is the case, use the following

instructions.

Fig.12.8 Schematic diagram showing the calibration task Set-Up