3 installing the 1314i in a 19” rack, 4 setting the usb communication parameters, 5 setting the rs-232 communication parameters – LumaSense Technologies INNOVA 1314i User Manual

Page 17: 1 default rs-232 communication parameters, Section 2.3, Section 2.4, Section 2.5

Advertising
background image

Chapter 2

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 17 of 199

2.3 Installing the 1314i in a 19” rack

When installing the 1314i in a 19” systems rack.

Place the Gas Monitor on a Shelve or support it by using Slide Rails.

Secure the Front frame to the 19” section with appropriate screws.

2.4 Setting the USB Communication Parameters

The LumaSoft Gas 7810 application with USB communication is able

to communicate with the 1314i gas monitor through a USB cable.

No setup on the 1314i gas monitor is necessary in order to be able to

communicate with the 1314i gas monitor through an USB connection.

2.5 Setting the RS-232 Communication Parameters

2.5.1

Default RS-232 Communication Parameters

The default RS232 communication parameters are shown in the table

below.

Baud rate

9600

Stop bits

1

Data bits

7

Parity

Even

Hardwire mode

Leased line

Handshake type

Hardwire

Text line Terminator CR-LF

Performing a full reset will set all parameters to the default settings.

2.5.2

Checking/Changing the RS-232 Communication

Parameters

The communication parameters for the serial interface must be set

using the push-buttons on the front of the monitor.

1. Press SET-UP S3 S1 S3 S1 .The screen display now shows the

following text.