LumaSense Technologies INNOVA 1314i User Manual

Page 52

Advertising
background image

Chapter 4

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 52 of 199

1. Press q and u.

The following text appears on the display:

WATER VAPOUR MEASURED

The displays which follow, indicate the type of text which appears

each time q is pressed:

WATER: MAX 8.55E+00 m 8.15+00

ALARM 1 MIN 7.01E+00 s 816E-03

This is the statistical report of the monitored gases/vapours

displayed when the monitoring task is complete. In this case where:

MAX

= maximum concentration of water vapour measured

during the whole monitoring task.

MIN

= minimum concentration of water measured during the

whole monitoring task.

ALARM 1 =

the user-defined concentration of water vapour

which, if measured by the Monitor, will trigger a switch

in the Monitor and activate an external alarm relay

connected to it.

m

= linear average, or mean value, of all the water vapour

concentrations measured during the whole monitoring

task.

s

= standard deviation of water-vapour’s concentrations

from the mean value (m) described above (see

section

8.2.1

for further details).

The humidity unit in this case is Tdew because this is the unit chosen

in the active set-up (see

section 4.3.2

).

W: WATER VAPOUR: 971.88mBar

2014-01-16 09:50:32 1 7.49Tdew

W: WATER VAPOUR: 971.88mBar

2014-01-16 09:51:33 2 7.49Tdew

As you can see the measurement cycles are numbered sequentially

as soon as the monitoring task has been stopped for example, the

above measurement was made at 09:51:33 on the 16

th

of January

and it was the second measurement cycle of the monitoring task.

The concentration of water vapour is given in the units of Tdew

because this was the unit chosen for humidity in the active set-up.