3 tasks before starting any calibration – LumaSense Technologies INNOVA 1314i User Manual

Page 137

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 137 of 199

Presence of Other Interferents

In many measurement situations water vapour is not likely to be the

only interferent present. Suppose that you have installed an optical

filter “A” to measure a gas (A), but another gas (B) is normally found

in the ambient air you are monitoring. If gas B absorbs some light

from filter “A”, it will “interfere” with the signal produced by gas A’s

absorption of this light. This is illustrated by the lower absorption

spectrum in Fig.12.3 (“Interference from Gas B when filter A is

used”). This interference is termed cross interference.

How is this interference measured? An optical filter “B” is chosen,

which transmits light with a wavelength which is most strongly

absorbed by gas B, and less strongly by gas A, this is illustrated by

the lower absorption spectrum shown in

Fig.12.3

. During cross

interference calibration two sets of measurements are performed:

1. The signal in the cell is measured using filter “A” and the filter “B”

when a known concentration of gas A is in the analysis cell. The

ratio of these two signals is directly related to gas A’s relative

absorption of light at these two wavelengths (upper spectrum in

Fig.12.3

). This ratio is a constant and is a measure of Gas A’s

Interference on filter “B”.

2. The signal in the cell is measured using filter “A” and filter “B”

when a known concentration of gas B is in the cell. The ratio of

these two signals is directly related to gas B’s relative absorption

of light at these two wavelengths (lower spectrum in

Fig.12.3

).

This ratio is a constant and is a measure of Gas B’s Interference

on filter “A”.

By measuring the ratios described above, the Monitor is able to

compensate any measurement made with filter “A” for any

interference signal produced by the presence of gas B in the cell, and

vice versa. This procedure is known as cross compensation.

12.3

Tasks Before Starting Any Calibration

Before starting any kind of calibration task there are several

operations, which need to be performed:

12.3.1 Obtaining a Print-out of Calibration Data in the

Monitor

Whether the Monitor is to be used online or as a stand-alone

instrument, the Calibration Software BZ7002 can be used to check

the Monitor’s existing calibration. Please refer to the User Manual

BE6034.

Advertising