Section 12.11 – LumaSense Technologies INNOVA 1314i User Manual

Page 176

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 176 of 199

12.10.4 Performing a Span Calibration of the Water-vapour

Filter

Section 12.4.1

describes a good method of producing water-vapour

of known concentration. The directions for performing a span

calibration of water-vapour filter are similar to those for performing a

humidity-interference calibration of the other installed filters is as

described in

section 12.7.2

. In both calibration tasks the signal in the

cell is measured when a supply of zero gas containing a constant

concentration of water vapour is attached to the Monitor. There are,

however, two differences:

1. During a humidity-interference calibration the signal is measured

using each of the installed filters (including the water-vapour

filter), whereas during the span calibration of the water-vapour

filter the signal is only measured with the water-vapour filter.

2. During a humidity-interference calibration, it is not necessary to

know the concentration of water-vapour supplied to the cell as it

is only necessary to measure the size of each signal (with filters

“A” – “E”) relative to the size of the signal measured with the

water-vapour filter. However, during a span calibration of the

water-vapour filter it is necessary to know the exact

concentration of water-vapour supplied to the cell, because during

this calibration the conversion factor is calculated. This factor

relates the measured signal to the concentration of water-vapour

in the cell.

Follow the procedure described in

section 12.7.2

.

12.11 Setting-up a Combined Zero-point and

Humidity-interference Calibration – Stand-

alone Use

We have explained in previous sections of this chapter (

section 12.6

and

section 12.7.1

) that a zero-point calibration can only be

performed alone if a humidity gain factor is already stored in each

of the filter banks which are active during the zero-point calibration

task. Likewise, a humidity-interference calibration can only be

performed alone if a concentration offset factor is already stored

in each of the filter banks which is active during the humidity-

interference calibration task.

If the necessary calibration factors are not stored in the Monitor’s

memory, a zero-point calibration task has to be performed together

with a humidity-interference calibration task.

Advertising