Chapter 11

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 128 of 199

Fig.11.5 Cleaning the surfaces of the internal air-filtration unit

Notes:

Only proceed to the next step when the acetone has completely

evaporated from the air-filter retaining disc and filtration unit.

Each of the fine filter-papers DS 0759 are packed between two

pieces of packing paper. The fine filter-paper is always white in

colour.

Remove a new fine filter-paper from its packaging using the

tweezers. Hold it by its edge. Refer to Fig.11.6. Lift the filter up and

place it inside the unit.

Fig.11.6 Placing a new filter-paper in the unit

Holding the handle-end of the unit between your thumb and fingers,

pick up the (dry) retaining disc and place it over the new filter-paper.

Refer to

Fig.11.7

. Gently press the disc down over the filter-paper,

making sure that the disc’s locking tabs fit properly into the groove

in the unit and that the filter-paper stays in position and is not

damaged in any way.

Holding the handle-end of the re-assembled unit between your

thumb and fingers, place it back into position in the air-inlet of the

Monitor. Screw it back into its mounting. To do this, turn the unit

clockwise with your fingers to tighten until it is in position.