1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 196 of 199

CLOCK SET TO DEFAULT ERROR

DETECTED IN CLOCK SETTINGS

The Monitor’s internal

clock was found to be

incorrectly set so these

values have been set to

their default values

A new battery has

been installed

FACTORS SET TO DEFAULT ERROR

DETECTED IN CALIBRATION

FACTORS

An error has been found in

the data stored in the

Calibration Factor part of

the Monitor’s memory, so

these factors have been

set to their default values

1. A defective

back-up battery

2. A defective

Working Memory

(RAM) or Source

Memory

(EEPROM)

3. A Software fault

MEMORY SET TO DEFAULT ERROR

DETECTED IN BACKGROUND MEMORY

An error has been detected

in the Monitor’s

Background Memory, so

the Monitor automatically

deletes all data stored in

this Memory

1. A defective

back-up battery

2. A defective

Working

memory ( RAM )

3. A software fault

MEMORY SET TO DEFAULT ERROR

DETECTED IN DISPLAY MEMORY

An error has been detected

in the Monitor’s Display

Memory, so the Monitor

automatically deletes all

data stored in this

memory.

1. A defective

back-up battery

2. A defective

Working Memory

( RAM )

3. A software fault

MEMORY SET TO DEFAULT ERROR

DETECTED IN INTERNAL MEMORY

An error has been found in

the data stored in the

Monitor’s internal memory.

data stored in this memory

cannot be read or altered

by the user, so the Monitor

automatically corrects any

errors found in this

memory

1. A defective

back-up battery

2. A defective

Working Memory

( RAM)

3. A software fault.

MEMORY SET TO DEFAULT ERROR

DETECTED IN SOURCE MEMORY

An error has been found in

the data stored in the

Monitor’s source memory,

so the Monitor

automatically gives stored

parameters default values.

1. A defective

back-up battery

2. A defective

Source Memory

( EEPROM)

3. A software fault

PARAMETERS SET TO DEFAULT

ERROR DETECTED IN

CONFIG:/FORMAT PARAMETER

An error has been found in

the value parameters

stored in the Configuration

and/or Format branches of

the Monitor’s set-up tree,

and therefore the Monitor

has automatically given

these parameters default

values

1. A defective

back-up battery.

2. A defective

Working Memory

(RAM)

3. A software fault.

PARAMETERS SET TO DEFAULT

ERROR DETECTED IN THE TASK

SET_UP PARAMETER

An error has been found in

the value parameter stored

in the Monitoring Task

branch of the Monitor’s

set-up tree, and has

automatically given these

parameters default values.

1. A defective

back-up battery

2. A defective

Working Memory

(RAM)

3. A software fault.