3 checking and changing optical filter parameters, Section 12.3.3, Fig.12.4 – LumaSense Technologies INNOVA 1314i User Manual

Page 141

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 141 of 199

Fig.12.4 Part of a calibration data print-out

12.3.3 Checking and Changing Optical Filter Parameters

Before starting calibration it is always wise to check that information

about the installed optical filters has been “entered” correctly in the

active set-up of the Monitor. If the incorrect UA number is

“entered” for the optical filter in a particular position, the incorrect

optical filter factors will be used in the calculation of the calibration

factors, and calibration will therefore be useless.

The UA number can only be changed using the push-buttons on the

front panel of the monitor.

1. Press SET-UP S3 S3.

The following text appears on the screen display:

Advertising