Simple test to determine the linearity – LumaSense Technologies INNOVA 1314i User Manual

Page 164

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 164 of 199

·

The High Concentration should be 75% to 100% of the highest

concentration you expect to measure, but no higher than this.

Simple Test to Determine the Linearity

Let us assume that you wish to measure a gas over a dynamic range

from 100 ppm and 3000 ppm of a gas that has a detection limit of

0.3 ppm.

In this case, the range in which you want to measure is from 333

times to 10 000 times the detection limit. With our general

guidelines, you’re not able to decide whether you have to perform a

single- or two-point calibration. Therefore, we advise you to perform

the following test:

1. Perform a single-point calibration with a low concentration of the

gas. Use 10% of the highest expected concentration level. In this

case, use 300 ppm (1000 times the detection limit).

2. Perform a measurement with a high concentration of the gas.

Use, for example, 90% of the highest expected concentration

level. In this case, use 2700 ppm.

If the measured gas concentration differs by more than 5% from the

real concentration, then a two-point calibration with 2 gas

concentrations is recommended.

In the example in

Fig.12.15b

, we have measured 2400 ppm with

2700 ppm gas. This gives an error of 11% and an even greater error

at the highest concentration level.

Advertising