3 setting-up the monitor’s alarm levels – LumaSense Technologies INNOVA 1314i User Manual

Page 84

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 84 of 199

23. The text screen returns to the top of the set-up tree. Press

SET-UP to exit the set-up tree.

If you want to set Alarm limits for the measured gases, go to section

7.3.

7.3 Setting-up the Monitor’s Alarm Levels

The alarm limits of a gas are those concentrations of gas which

should activate any alarm relay connected to the Monitor via its back

panel “Alarm Relay” socket. There are two different alarm relays,

which function slightly differently. Details of these relays and the way

in which they function is described fully in

section 5.5

. The alarm

level parameters are found in the Filters branch of the Set-Up Tree.

Warning! If the UA number for any filter is changed here, then all

the calibration data for that filter will be lost and the filter will need

to be recalibrated.

1. Press SET-UP S3 S3. The following text is displayed

CHECK AND/OR CHANGE SET-UP FOR FILTER A

NO YES

2. Use the select push-buttons to ignore this filter or to select this

filter if alarms are to be set.

3. Press q 5 times until the following text is displayed.

GAS A HIGH ALARM LIMIT 1 _________mg/m

3

PRESS ENTER TO CHANGE VALUE

4. Use the direction push-buttons to set the alarm concentration.

If you do not wish to activate the alarm function for a particular

gas, select the DEFAULT. This will “blank out” the alarm level

value for the gas and deactivate the alarm for that gas.

GAS A HIGH ALARM LIMIT 2 _________mg/m

3

PRESS ENTER TO CHANGE VALUE

5. Repeat steps 2 to 4 to set the alarms for the other filters.

6. Press SET-UP to exit the alarm set-up.