Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 83 of 199

Environment branch

13. Press S3 to enter the Environment branch.

14. Use the select push-buttons to select the desired flushing

mode.

If you select Auto, go to step 17.

If you select Fixed Time, go to step 15.

15. Use the direction push-buttons to define the Chamber Flushing

Time.

16. Use the select push-buttons to select Tube Flushing.

If you don’t want tube flushing, select NO and go to step 18.

If you select tube flushing, use the direction push-buttons to

define the flushing time and go to step 18.

17. Use the direction push-buttons to select the desired sample

tube length and move on to the next display.

18. Use the direction push-buttons to select the desired

normalization temperature.

See section 7.1

for details about the normalization options.

When step 18 is completed, the display returns to the top of the

MEASUREMENT branch.

Press p to go to the top of the Set-up Tree.

Press SET-UP to exit the set-up function.

Format branch

If you are starting here, press SET-UP:

19. Press S2 to enter the FORMAT branch.

20. Use the select push-buttons to choose if you want to print each

gas.

If you select Yes, go to step 22.

If you select No, go to step 23.

21. Use the select push-buttons to choose which individual

parameters should be printed. Each time one of the select

push-buttons is pressed, the next parameter is displayed. This

step can be repeated up to 7 times, depending on the number

of filters installed.

22. Use the select push-buttons to select the Text line Terminator.