Chapter 8

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 97 of 199

8.3 Storage Space in the Display Memory

The total space available in the Display Memory to store data is

131072 measurement cycles. Depending on the time for one

measurement cycle the capacity will be as follows.

If a measurement cycle in which gas A and water vapour are

measured with continuous sampling takes 15s, then the Display

memory space will be 546 hours = 22 days.

If a measurement cycle where 5 gases and water vapour are

measured with continuous sampling takes 60 s, then the Display

memory space will be 2184 hours = 91 days.

The Monitor will overwrite the oldest stored measurement data when

the Display memory is full. This is done by deleting data in blocks of

2048 measurement cycles.

For further information regarding measurement cycles see