1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 126 of 199

Whenever the Monitor needs to be calibrated, we recommend that

the fine filter-paper in the internal air-filtration unit is changed

before you calibrate. The fine filter-paper in the internal air-filtration

unit must be changed at least twice a year; and more frequently if

the environment in which the monitor is working contains large

quantities of particulate material suspended in the air.

When the monitor is making measurements, we recommend that a

length of Teflon

®

tubing (of up to 50 metres in length) is always

attached to the stub of the air-inlet (see

Fig.11.1

) and that the

special external air-filtration unit UD 5023 (optional accessory),

shown in Fig.11.3, is always attached to the free-end of the sampling

tube, by means of the small stub. An external coarse metal air-filter

prevents large particles, insects etc. being drawn into the unit, while

inside the unit is a fine (10µm) filter-paper, which prevents any finer

particles from entering the Teflon

®

sampling tube. The fine filter-

paper in this unit must also be changed regularly – at least as

frequently as the fine filter-paper in the internal fine air-filter.

Fig.11.3 The external air-filtration unit UD5023 used with the

Monitor, (optional accessory)

Changing the fine filter-paper in both the internal and external air-

filtration units is explained step by step in the following two sub-

sections.

11.1.1 Changing the Filter-paper in the Internal Air-

filtration Unit

While you do this, we recommend that you wear clean, rubber

gloves.

Tools and equipment required: