Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 154 of 199

RESULTS NOT YET AVAILABLE

PLEASE WAIT

Once the first measurement result is available the screen text

illustrated above is replaced by a screen displaying the

measurement results – for example the following:

W: µ : 11.3µV σ --- V 1 38.6°C

PRESS ENTER WHEN RESULTS ARE STABLE

As soon as measurement results are available they not only

appear on the display (see Fig.12.12) but are automatically

printed out on the printer, if the Monitor has been set up to print a

data log, see

section 10.5

. Remember to turn off the data log

when the calibration is complete.

Fig.12.12 Example of data displayed during a calibration (the text

below the display describes the displayed data)

Each time a sample of gas is drawn into the analysis cell the

signal is measured using all the installed optical filters and the

water-vapour filter (“W”) but, due to lack of space on the display,

the Monitor only displays the signal measured using the water-

vapour filter. You cannot scroll through the measurements. The

data-log print-out, however, shows the signal measured using all

installed filters as well as the water-vapour filter.

A maximum of 6 measurements are stored in the Monitor during

any calibration. The number (n) indicates how many

measurements are stored. When 7 measurements have been

performed the number (n) will show the number 6 because the

very first measurement is overwritten by the 7

th

measurement so

that only 6 measurements are still stored. When 8 measurements