water-vapour filter must be stored in the Monitor’s calibration data

block.

During span calibration of the water-vapour filter it is necessary to

attach a supply of zero-gas containing a known concentration of

water vapour to the Monitor. As explained in the Humidity-

interference Calibration (

Section 12.7.2

) it is important to use water-

vapour whose concentration is below the saturated vapour pressure

of water-vapour at the temperature of the room where calibration is

to be performed, otherwise water-vapour will condense out in the

analysis cell and damage it.

Before setting-up the span calibration of the water-vapour filter go

into Set-Up mode and select “Tdew” as a humidity unit. Then when

you have to enter the concentration of the water-vapour you will use

during calibration. Always use water concentrations at least 2°C

below the ambient room temperature. For example, if the ambient

room temperature is 20°C then you can ENTER WATER-VAPOUR

CONC as 18°Tdew.

The procedure for setting-up a span calibration of the water-vapour

filter is as follows:

1. Press Measure S3 S3 S1 then the following text will be

displayed:

PERFORM ZERO-POINT CALIBRATION ?

NO YES

2. Follow the set-up of a span calibration of the water-vapour filter

by following the black arrows in the schematic diagram shown

in

Fig.12.20

.

The following text will appear:

CONNECT WATER-VAPOUR 18.0 Tdew

AND PRESS ENTER WHEN READY