1 measurement set-up – stand-alone use, Table 7.6 – LumaSense Technologies INNOVA 1314i User Manual

Page 80

Advertising
background image

Chapter 7

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 80 of 199

Table 7.6 The range of acceptable values and default value of the alarm

limit for the gases

Display Text

Acceptable Values

Default

Value

GAS A HIGH ALARM LIMIT 1

n

à

_____PPM

0-1000000ppm

Blank

GAS A HIGH ALARM LIMIT 1

n

à

_____mg/m3

Depends on the molecular

weight of the gas

Blank

n

à

= A to W

Action: this provides you with 3 ways to trigger the alarm. One or

more of the options can be selected simultaneously.

Monitor Beep: when an alarm is triggered the monitor makes a

beeping noise.

Monitor Relay 1 & Monitor Relay 2: selecting these options

enables you to connect the alarms to a variety of either audio

(for example, a siren) and/or visual (for example a flashing

light) alarm systems.

Active Alarms: only the gases, which you have selected and

activated, are shown in this field.

7.2.1

Measurement Set-up – Stand-alone Use

The philosophy of the instructions is described in

Chapter 3

.

To enter the Set-Up Tree, press SET-UP and the monitor displays

the following screen text.

1.

Press S1 to enter the MEASUREMENT branch. The new text

displays two branches.

SELECT SET_UP BRANCH

MEASUREMENT FORMAT CONFIGURATION