14.3

“Invalid” Calibration Messages

The messages listed below indicate: (1) that the calibration factor

calculated during the calibration task you have just performed is not

valid and therefore has not been stored in any filter bank; and (2) that

the calibration task must therefore be repeated.

ZERO-POINT CALIBRATION INVALID*

HUM. INTERFERENCE CALIBRATION INVALID*

ZERO-POINT/HUM. INTERF. CALIB. INVALID*

GAS X* SINGLE-POINT CALIB. INVALID*

GAS X* TWO-POINT CALIBRATION INVALID*

14.4

Error Messages Displayed when the INFO

button is used

All “successful*” or “invalid” calibration messages are marked with an

asterisk (*) which tells you that more detailed information about the

calibration can be obtained by pressing INFO. The messages

displayed when INFO is pressed are listed and explained below.

14.4.1 “Invalid” Error Messages

MISSING MEASUREMENT DATA

This message indicates that the Monitor does not have the necessary

data to calculate absolutely valid calibration factors. This situation

could, for example, be caused by too much noise in the analysis

chamber. The calibration task will have to be repeated.

WATER VAPOUR MEASUREMENT < OFFSET

This message is associated with either (1) a span calibration of the

water-vapour filter, or (2) a humidity-interference calibration.

This message indicates that the signal measured with a particular

filter, when water-vapour of known concentration is in the analysis

cell, is less than the signal measured in the cell when zero-gas is in

the cell. This could indicate either:

·

That the gas used during the zero-point calibration of the filter was

not dry;

or

·

That the water vapour used during either the span calibration of the

water-vapour filter; or the humidity-interference calibration, was

not wet enough.