LumaSense Technologies INNOVA 1314i User Manual

Page 173

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 173 of 199

12.10.2 Performing a Zero-point Calibration of the Water-

vapour Filter

As explained in the previous section, a zero-point calibration of the

water-vapour filter has to be done together with the zero-point

calibration of the other installed optical filters. The latter task has

been fully described in

section 12.6

. It is worth-while to emphasise

again that it is vitally important to use perfectly dry zero-gas (that

is, zero-gas which contains 0% water-vapour) during this calibration.

Note that during the zero-point calibration of filters “A” to “E” (task

1, shown below), and during the combined zero-point calibration of

filters “A” to “E” and the water-vapour filter (task 2, shown below),

a zero-gas supply is attached to the air-inlet of the Monitor. The

Monitor measures the signal in the cell using the water-vapour filter

and all the other installed optical filters. Signals measured with filters

“A” to “E” enable the concentration offset factors for filters “A” to

“E” to be stored in the filter banks which are active during

calibration. However, the signal measured with the water-vapour

filter is used differently:

In Task 1:

The signal measured with the water-vapour filter is used to

compensate all measured signals for any interference caused by the

presence of any residual water-vapour in the zero-gas supply.

In Task 2:

The signal measured with the water-vapour filter is used to update

the concentration offset factor for the water-vapour filter. This

means that the signals measured with filters “A” to “E” cannot be

compensated for the presence of any water vapour present. This is

why it is vitally important to use perfectly dry zero-gas during

this task.

Please refer to

section 12.6.2

for details about how to perform a

zero-point calibration task.

12.10.3 Setting-up a Span Calibration of the Water-vapour

Filter

Remember that span calibration of the water-vapour filter is only

necessary if you wish to measure the absolute concentration of water

vapour in air samples. If the water-vapour filter is not span

calibrated you will not be able to measure the concentration of water

vapour in gas samples during a monitoring task.

Before the water-vapour filter can be span calibrated, it has to be

zero-point calibrated, that is, a concentration offset factor for the

Advertising