Warning and error messages – LumaSense Technologies INNOVA 1314i User Manual

Page 195

Advertising
background image

Chapter 15

______________________________________________________________________

_____________________________________________________________________________

BE6030-12

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 195 of 199

Warning and Error Messages

The following table gives most of the error and warning messages

contained in the test Hierarchy of the Monitor. There is a description of

each fault as well as the possible cause of each fault. The user must

evaluate the significance of each message.

Warning Messages

Description of Fault

Possible Cause

AIR FLOW TOO LOW

The air pressure difference

created by the pump in the

pneumatic (airway) system

(tubes and/or analysis cell)

is too low. This means that

the tubing and/or analysis

cell cannot be properly

flushed out and the sample

in the analysis cell is

therefore not necessarily

“new”

1. The length of

the sampling

tube attached to

the Monitor’s

inlet has been

incorrectly

entered.

2. Either the

external or

internal

pneumatic

system is not

air-tight

3. The pump is

defective

4. This might also

indicate a too

high inlet

pressure.

AIRWAY-SYSTEM BLOCKED

The air pressure difference

created by the pump in the

pneumatic ( airway)

system ( tubes and/or

analysis cell) is too high

1. The Air inlet is

blocked.

2. Air Filter DS0759

blocked.

ALARM LIMIT EXCEEDED

The Monitor has measured

a concentration of gas

greater than the alarm

limit concentration entered

NO INDEX MARK FROM FILTER

CAROUSEL

The Monitor cannot

determine the position of

its filter carousel

1. The motor that

drives the filter

carousel is not

working

2. The optical

detector that

checks the

movement of

the carousel is

not working

3. The filter

carousel has not

been properly

mounted