Perform zero-point calibration? no yes, Connect, And press enter when ready – LumaSense Technologies INNOVA 1314i User Manual

Page 167

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 167 of 199

filters. Note that cross-interference calibration cannot be performed

alone, it has to be performed together with span calibration.

1. For each filter being span calibrated, enter the No. of the filter

bank where you wish the Monitor to store the conversion

factors which will be calculated during the span calibration task.

2. Press MEASURE S3 S3 S1 then the following text will be

displayed:

PERFORM ZERO-POINT CALIBRATION?

NO YES

3.

a. If you only wish to span calibrate filters, follow the set-up of a

span calibration task by following the black arrows in the

schematic diagram shown in

Fig.12.16

.

b. If you wish to perform cross-interference calibration as well as

span calibration, follow the set-up of the combined calibration

task by following the black arrows in the schematic diagram

shown in

Fig.12.17

.

The following text will appear on the display:

CONNECT

(the name of the gas appears here)

XXXXXXmg/m

3

AND PRESS ENTER WHEN READY

Note: if the gas concentration on the “Analysis Certificate” is given in

ppm then the following formula can be used to convert from parts

per million units (ppm) to mg/m

3

:

For a gas at 20°C and at 1 atmosphere pressure:

Concentration (in mg/m

3

) = Concentration (in ppm) x Molecular Weight (in g)

24,04