Chapter 9

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 106 of 199

9.1.5

Effect of Changing the UA number in the “Active”

Set-up

WARNING!: Changing the UA number of a filter in the Monitor’s

active set-up will destroy all calibration data for the filter.

As explained in

section 9.1.1

, the UA No. determines the set of

Optical Filter Factors used during the calculation of calibration factors

and therefore, if the UA number of any filter is changed in the active

set-up of the Monitor, all calibration data related to the optical filter

whose number has been removed from the set-up is automatically

deleted from both the Source Memory and the Working Memory

as it is no longer valid.

Fig.9.3 Effect of changing a UA number in the “active” set-up

For example, if the UA number of the filter installed in position “A” of

the filter carousel is changed in the active set-up then calibration

factors are deleted from the active file “A” in the Working Memory

and from files A 1, A 2, A 3, A 4 and A 5 in the Source Memory of

the monitor (see Fig.9.3). The Monitor will not be able to measure

any gas with filter “A” until it has been fully calibrated, see

Chapter

12

for details.