“W”

This symbol indicates that a warning was detected

during this measurement. The accuracy of such a

measurement cannot always be guaranteed.

Gas Marks:

“B”

The Monitor was incapable of calculating this gas

concentration.

“F”

After this measurement the filter carousel was

found to be out of alignment. If the carousel was

only slightly out of alignment then the accuracy of

the measurement has not been affected. The ac-

curacy of such a measurement cannot always be

guaranteed.

“A”

The gas concentration measured during this

measurement cycle was found to exceed the user-

defined alarm limit for this gas.

8.2.3

Symbols Used in Print-outs of Measurement Data

Displayed measurements marked with the COMMON MARKS “O”

and “W”, and the GAS MARKS “A” and ”F”, are printed-out with an

“X” symbol. However, any displayed measurements marked with the

COMMON MARKS “P” and the GAS MARK “B” are not printed-out

with these symbols. These measurements are marked in a different

way:

Common Marks:

The symbol “P” is replaced by the text “1314i Restarted”. This

indicates that the Monitor stopped and then started-up again

sometime between the previous sample time and the time which

appears alongside this message.

Gas Marks:

The symbol “B” is replaced by a straight line (_______) which

indicates that the Monitor was unable to calculate the concentration

of the named gas in this sample.

Other Marks Used:

……This symbol indicates either: (1) that this gas was not measured;

or (2) that this gas was not selected to be printed out (see