Fig.12.14 Curve showing a linear span-calibration curve

Choosing a filter bank

If a filter (e.g. “A”) has already been calibrated to measure a gas

(e.g. A1) and you wish to calibrate with gas A1 again, then select the

filter bank containing the gas conversion factor for gas A1. Online

users can do this when downloading the calibration factors. Stand-

alone users should do this before setting-up and performing the span

calibration task. The gas conversion factor calculated during the

span calibration will then just overwrite the gas conversion factor

calculated during the previous span calibration of the filter with this

gas.

If you wish to span calibrate a filter to measure more than one gas,

then select a new filter bank each time you span calibrate with a

different gas. In this way the conversion factor for each gas is

stored in separate filter banks (for stand-alone use see

Fig.12.9

).

The gas conversion factor stored in the selected filter bank during

a span calibration task will be overwritten by the new gas

conversion factor calculated during the span calibration of the

filter.

Gas to be Used During Span Calibration

The gas you need to use during span calibration must be the gas you

wish to measure with the filter.