LumaSense Technologies INNOVA 1314i User Manual

Page 16

Advertising
background image

Chapter 2

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 16 of 199

3.

Use a screwdriver to remove the fuse-holder:

4.

Remove the old fuse from the holder and replace it with a new.

Check that the fuse has the correct rating.

The fuse must be:

Slow-blow (T) fuse with a rating of 2.5 A, 250 VAC, 5x20 mm, UL

approved.

LumaSense order number: VF0102A

5.

Install the fuse-holder and re-assemble the 1314i again.