interference from water vapour. This demands that the zero-gas you

use during the zero-point calibration of the filter must be perfectly

dry.

12.10.1 Setting-up a Zero-point Calibration of the Water-

vapour Filter

If you wish to zero-point calibrate the water-vapour filter you must

zero-point calibrate all the other installed optical filters. The

concentration offset factors calculated during zero-point

calibration of the filter “A” to “E” will be stored in the filter banks,

(those active during calibration). Check, therefore, that you make

the relevant filter banks active for these filters before calibrating

them. This is discussed in more detail in

section 12.3.3.

The procedure for setting-up a zero-point calibration of the water-

vapour filter is as follows:

1. Press Measure S3 S3 S1 then the following text will be

displayed:

PERFORM ZERO-POINT CALIBRATION ?

NO YES

2. Follow the set-up of a zero-point calibration of the water-vapour

filter by following the black arrows in the schematic diagram

shown in

Fig.12.19

.

The following text will appear:

ZERO-POINT CALIB. CONNECT CLEAN, WET AIR

AND PRESS ENTER WHEN READY