Presence of water vapour – LumaSense Technologies INNOVA 1314i User Manual

Page 135

Advertising
background image

Chapter 12

______________________________________________________________________

_____________________________________________________________________________

BE6030-13

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 135 of 199

As it is the optical filter, which determines the wavelength of light,

cell noise will depend upon the optical filter being used.

The cell-noise signal is measured during the Zero-point Calibration

of each filter. A supply of dry, zero gas is attached to the air-inlet of

the Monitor and the signal in the cell is measured with each installed

optical filter (“A” to “E”). This signal is called the concentration

offset factor for the filter. Whenever the filter is used, this signal is

subtracted from the total measured signal as it is not related to the

concentration of any gas.

Presence of Water Vapour

Water vapour is nearly always present in ambient air and it absorbs

infra-red light, to a greater or lesser extent, at nearly all wavelengths

(see

Fig.12.1

). This means that no matter which optical filter is

transmitting light into the cell the water vapour in the cell will absorb

some of this light and create a signal. This signal is said to “interfere”

with the signal produced by the gas we wish to measure with this

filter.

Fig.12.2 Schematic diagram showing the relative absorption of water

vapour by different optical filters

How is this interference measured? A water-vapour optical filter,

which transmits light with a wavelength, which is absorbed by very

few gases, except water vapour, is always installed in position “W” of

the filter carousel. Using this filter, water-vapour’s interference is

measured during the Humidity Interference Calibration of each

installed filter. A supply of zero gas containing a constant

concentration of water vapour is attached to the air-inlet of the

Advertising